CATLAB-PCS.
A bench-top microreactor and combined mass spectrometer system for rapid and reproducible catalyst characterization and reaction studies.
Key Features.
- Single integrated software package allowing manual or automated control over both the microreactor and mass spectrometer parameters
- Quartz catalyst cartridge system for reproducible sample positioning
- Low thermal mass furnace for rapid linear response
- Sample in-bed thermocouple for optimum temperature accuracy
- Precision mass flow controllers for accurate flow measurement
- Zero dead volume valves ensuring rapid, reproducible response
- Pulse chemisorption option for uptake measurements, adsorption isotherms and catalyst dispersion
- Fully programmable and automated analysis cycle
Specifications.
- Analyzer
- Hiden quadrupole mass spectrometer system, which can also be used as a stand-alone instrument
- Mass range options
- Standard 0 to 200 amu, with options extending this to 1000 amu
- Detection limit
- 0.1 to 1 ppm, subject to spectral interference
- Ionization
- Soft ionization mode available for simplified spectra of complex molecules
- Standard inlet
- Heated inlet for sampling of condensable gases
- Response time
- Less than 500 ms to changes in gas concentrations
- Format
- Bench-top
- Control software
- CATLAB software
- System modules
- Module 1 is the microreactor including temperature and flow control. Module 2 is the Hiden quadrupole mass spectrometer system
- Data acquisition speed
- More than 500 measurements/s
Downloads.
Hiden's published documents for this instrument.
Related instruments.
CATLAB software
Software for the Hiden CATLAB systems, controlling MS data acquisition alongside devices such as MFCs, furnace temperature and switching valves.
SpaciMS
A spatially resolved capillary inlet mass spectrometer that determines species and temperature profiles, radially and axially, with high spatial and temporal resolution.
HPR-20 TMS
A transient mass spectrometer configured for fast event analysis of gases and vapors at pressures near atmosphere, ideal for fast gas switching experiments.
Talk to an applications engineer.
Describe the experiment. An applications engineer reads it personally and typically replies within one business day.